System Calibration And Setup




 
 

> click to enlarge <

System Calibration

The system must be calibrated in order to perform correctly. To do this, from the menu click Setup > Calibration: the window illustrated  will appear. By means of the " Load picture " button, load a previously scanned image, saved in a directory   according to your system organisation, with two points ten centimetres apart, use this to replace the example, follow the instructions, save and exit. The angles and relationships will always be correct. However, the distances will be  the same as the real ones, AND CONSEQUENTLY ALSO THE DIAGNOSTIC REPORT, only when the system has been correctly calibrated.

If the calibration configuration is modified following modifications to the operating system (for example, if you decide to scan with a different resolution from the one currently preset), this will affect the linear values of all previously elaborated images.

It is advisable to carry out some trials first in order to adapt results to the hardware currently being used, in order to set up an operating standard that will not require any further modification.

The bitmap images in the programme files refer to x-rays scanned at 75 dpi. The above programme default coefficients  regard images imported from a scanner, and a 17" monitor for screen processed images.

This means that if a 75 dpi resolution image is used when scanning images for later acquisition, or if the image to be used is imported directly from a 17" monitor screen, the system doesn't have to be calibrated.

If the image captured at 75 dpi (recommended resolution) is satisfactory, and the screen being used is not a 17" one, when calibrating the system for the first two points, use the default image, whereas for the second two points, place a ruler above the screen and click at a distance of 10cm.

Given that even when a 17" monitor is used, minute differences can be found depending on the particular regulation that is set when the monitor is being adjusted, it is advisable to carry out the calibration to further ensure system accuracy.

To make these procedures as simple as possible:

always use 75 dpi to capture bitmap images and carry out system calibration as per the instructions using the previously loaded image.

 

[Back To Index]